GEA WITH MK2.1 AND L-PROBE

  • Oxygen measurement in gas atmospheres
  • simple gas sampling
  • Inexpensive
  • Reliable
  • Insensitive to temperature
  • Insensitive to vibration
  • Easily replaceable
  • Fitting adapted to customers needs

The probe measures the residual oxygen content in a measuring chamber at the end of the measuring gas fitting.

The probe signal can be tapped off as a voltage. The probe is heated with a constant voltage. For supplying the heating only power supply inits approved by MESA may be used.

The quantity of gas passing the probe should be within a range of 20…50 l/h. Too much gas volume can change the highly constant probe temperature, which results in a falsification of the measured value. Every probe includes two constants (K1 and K2). These constants must be entered in the measuring instruments (e.g. intelligent measuring transmitter Carbo47,Carbomat-M, MCON Carbo).Ifnottherewillbe some measuring faults.

Dimension of sensing head:

105 x 290 x 45mm (BxHxD)

Rated length:

400…1100 mm

Connection L-probe:

snap closure

Necessary additional parts:

NTV44P for heating the L-probe and controlling temperature.

Measuring range:

0…1300 mV

Optional devices

Carbo 15/ 47
MCon Carbo 16-48
Carbo 100
Carbo 1000
Carbomat-M
MCON Carbo

OUR PRODUCT IN THE INDUSTRY

This unit is designed for use in the heat treatment industry