GEA-MK2.1 AND L-PROBE WITH CERAMIC VERSION (CHILLER AS AN OPTION)
- Oxygen measurement in gas atmospheres
- simple gas sampling
- Inexpensive
- Reliable
- Insensitive to temperature
- Insensitive to vibration
- Easily replaceable
- Fitting adapted to customers needs
The probe measures the residual oxygen content in a measuring chamber at the end of the measuring gas fitting. The probe signal can be tapped off as a voltage. The probe is heated with a constant voltage. For heating and supplying only power supply approved by MESA may be used.
The quantity of gas passing the probe should be within a range of 20…50 l/h. Too much gas volume can change the highly constant probe temperature, which results in a falsification of the measured value. Every probe includes two constants (K1 and K2). These constants must be entered in the measuring instruments (e.g. intelligent measuring transmitter Carbo48, Carbomat-M, MCON Carbo), if not there will be some measuring faults.
Dimension of sensing head:
105 x 290 x 45mm (BxHxD)Rated length:
400…1100 mmConnection L-probe:
snap closureNecessary additional parts:
NTV44P for heating the L-probe and controlling temperature.Measuring range:
0…1300 mVOptional devices
Carbo 15/ 47MCon Carbo 16-48
Carbo 100
Carbo 1000
Carbomat-M
MCON Carbo
OUR PRODUCT IN THE INDUSTRY
This unit is designed for use in the heat treatment industry