DEWCHECKER 2.1
ADVANTAGES:
- Easy handling
- High measuring certainty and accuracy
- Insensitivity to gases containing dust and corrosive gases
- Robust construction
- Safe designation of dew point by detaining the temperature and finevon +/- 0,2°C adjustments in steps of +/-0,2°C
Design
Portable steel housingDimensions
330x310x320 mm (WxHxD)Weight:
Approx. 17 kgDegree of Protection:
Housing IP 20 according to IEC 60529Front panel IP 40 according to IEC 60529
Power supply
230 or 115 VAC ±10% 50…60 Hz (please specify)Measuring Range:
-25…+30°C (direct measurement with mirror)-80…+20°C (indirect measurement with capacitive sensor)
Measuring Gas Connectors (hose connectors):
Input: Tube outer diameter 5…6 mmOutput: Tube outer diameter 5…6 mm
Built-in measuring pump
Measuring Gas Pressure:
±50 mbarMeasuring Accuracy:
±0,5°C (direct measurement with mirror)measurement with capacitive sensor)
Analog Output:
-80…+20°C, 0…20 mA or 4…20 mA or 0…10 V himiditygalvanic isolated, 4…20 mA is the predefined setting
Humidity
Up to 70 %, non-condensingAmbient Temperature:
10…40 °C during operation0…50 °C during storage
Gas Flow:
Approx. 1l/min.Resolution in steps of:
0,1°CMaintenance Interval:
Every yearDigital Output:
Either RS-485 / RS-422 four-wire or RS-485 two-wire (please specify)Isolated interface with Modbus-protocol
EQUIPMENT
Power cord
product code 011-0520Polishing cotton
product code 130-6050Polishing paste Connection pipe yellow
No product code availableConnection pipe yellow
product code 015-71038 mm with adapter approx. 3 m
Connection pipe yellow p
roduct code 015-71026 mm approx. 3 m
In-line filter
product code 24382Filter material
product code 24221Wooden stick for cleaning
no product code availableSpare fuse (230V)
product code 027-0005Spare fuse (115V)
product code 027-0011Spare sinter filter insert
product code 2418