Foil testing set FPG 1.1

  • Number of slides is set by the user
  • Fast and reliable measurement
  • New algorithm detects mechanical vibrations and compensates for them
  • The user can choose between 5 “filter algorithms” in order to use the foil testing device in a “laboratory or harsh environment”
  • Measurement accuracy of 0.01 [mg]
  • Direct display in% C and [mg]
  • Portable analysis system
  • Clear text operator guidance on the display
  • Foil scale can be operated without a PC or with a PC connection

PRINCIPLE OF OPERATION:

With the foil tester FPG 1.1, you can immediately determine and display the carbon content in furnace atmospheres with the help of test foils. Additional calculations that show the difference in the weight of the foils are not necessary! Each step is clearly shown on the terminal display so that the device can be operated without errors

SPECIAL FEATURES

 

  • Fast, problem-free determination of actual carbon level values ​​in furnace atmospheres
  • Simple operation via plain text menu navigation in the following languages: German, English, Italian, French, Spanish, Portuguese and Czech
  • Automatic and quick display of the percentage by weight of carbon in the furnace atmosphere in% C
  • Different foils can be tested at the same time
  • Automatic calibration cycle
  • Storage of the data
  • Complete with basic accessories

Design:

Table case

Weight:

10.25 kg

Auxiliary voltage:

85-264 VAC – 50-60 Hz

Measuring accuracy:

0.01 [mg]

Measuring range:

90 [mg] to 120 [mg] 0.01 … 1.99% C

Foil base weight:

In% C

Service:

Menu navigation via terminal

Dimensions:

200x225x395 mm (WxHxD)

Protection class:

IP 54 according to DIN 40050

Power consumption:

Max. 15 W

Resolution:

0.001% C

Climate conditions:

10 … 40 ° C
5 … 95% relative humidity, free of condensation

Display:

Display with 4×16 characters

Duration of the film analysis:

Approx. 15 s

EQUIPMENT

Calibration weight 95mg with case

100-0501

Test foils

630-0202

Antistatic Tweezers

370-1010

Round wood

100-0601

Acetone container

380-0010

OUR PRODUCT IN THE INDUSTRY

This unit is designed for use in the heat treatment industry