DEWCHECKER 0.1

ADVANTAGES: 

  • Easy handling
  • High measuring certainty and
  • Insensitivity to gases containing dust
  • Robuste construction

DEWCHECKER 0.1

FEATURES:

The dew point measurement instrument Dewchecker 0.1 is a portable device used to measure the dew point of gases. Per definition, the dew point is the temperature of a body marking the start of condensation of water vapor on its surface respectively the balance adjusted between condensing and evaporating water.
The gas sampling occurs via a gas sampling fitting furnished in the hot zone to the furnace wall with a ceramic inner tube.

Due to a steel tube length of at least 150 mm, outside the furnace wall, the sample gas is cooled quickly and guided at a gas dew point temperature below room temperature through a dense gas hose to the gas inlet port of the device.
The Dewchecker 0.1 is equipped with a capacitive on-line sensor used accordingly for continuous measurements. The measurement result can be forwarded by a universal analog output or a digital RS-485 / RS-422 interface.

Design:

Portable steel housing

Dimensions:

330x310x320 mm (WxHxD)

Weight:

Approx. 17 kg

Degree of Protection:

Housing IP 20 according to IEC 60529
Front panel IP 40 according to IEC 60529

Ambient Temperature:

10…40°C in during operation
0…50°C during storage

Power Supply:

230 or 115 VAC ±10% 50…60 Hz (please specify)

Measuring Gas Connectors (hose connectors):

Input: Tube outer diameter 5…6 mm
Output: Tube outer diameter 5…6 mm
Built-in measuring pump

Messgasdruck:

±50 mbar

Measuring Range:

-80…+20°C (indirect measurement with capacitive sensor)

Measuring Accuracy:

±2°C (indirect measurement with capacitive sensor)

Resolution in steps of::

0,1°C

Maintenance Interval:

Every year

Analog Output:

-80…+20°C, 0…20 mA or 4…20 mA or 0…10 V galvanic isolated, 4…20 mA is the predefined setting

Pin Assignment:

Current output: Pin 6, GND = Pin 2
Voltage output: Pin 7, GND = Pin 3

Digital Output:

Either RS-485 / RS-422 four-wire or RS-485 two-wire (please specify) Isolated interface with Modbus-protocol

Pin Assignment:

Four-wire: GND = Pin 1, TxA = 9, TxB = 5, RxA = 8, RxB = 4
Two-wire: GND = Pin 1, A = 9, B = 5

EQUIPMENT

Power cord

product code 011-0520

Connection pipe yellow

product code 015-7103

8 mm with adapter approx. 3 m

/

Connection pipe yellow

product code 015-7102

6 mm approx. 3 m

product code 24382

Filter material

product code 24221

Spare fuse (230V)

product code 027-0005

Spare fuse (115V)

product code 027-0011

Spare sinter filter insert